§ For use with ILT1700 to monitor and record data from 1 to 8 ILT detectors(example system configurations by application listed below)
§ Rapidly change between detectors via software (included) or manually using the active channel buttons
§ Data log up to 2 billion data points with time and date stamp
§ 10 Decade Dynamic Range
§ Programmable units of measurement include lux, W/cm2, Watts, candela, many others
§ Remote control of many ILT1700 features including Hold, Zero, DC and Integrate
§ Software displays light measurements in real-time
§ Export data into most standard analysis software such as Microsoft Excel®
§ National Instruments LabView® interface/driver that allows customization is available |
|
The ILT Model A803 is designed to provide the ILT1700 Research Radiometer with the ability to continuously monitor and collect light measurement data from up to eight (8) ILT detectors. Many combinations of ILT detectors up to eight total may be used and can be remotely mounted with custom cable lengths up to 100 feet in length.
Use of the A803 with the ILT1700 Research Radiometer eliminates the need to manually change detectors and calibration factors. The A803 software records the channel and calibration factor and allows the researcher to quickly change between detectors without having to unplug and exchange detectors on the rear panel of the ILT1700. This reduces the risk of using the wrong calibration factor and decreases the set-up time. It also allows user control of functionality of the ILT1700 including, zero, hold, integrate and DC mode selection.
A803-8 Front Panel
§ Function control panel indicates ILT1700 features zero, dc, integrate and hold
§ Active Channel buttons manually switch between detectors
§ LED channel indicators show which detector is in use |
|
A803-8 Back Panel
§ 8 detector ports for sensor connection
§ USB port connects to computer
§ Accessory input allows control of ILT1700 functions |
|
Configuration Information:
Optimum system configuration: A803, ILT1700, personal computer (laptop or desktop) and up to 8 ILT detectors.
Two Model ILT A803 configurations are available:
A803-8 (8 detector channels)
A803-4 (4 channels)
Specifications:
Input: Up to 8 ILT detectors Output: USB 2.0 Dimensions: 2.625 x 7.5 x 4.625 Weight: 1.75 lbs Data Acquisition Rate: 0.5 seconds to 12 hours Computer Requirements: Operating System: Windows XP, Vista, 7 Data Output: Excel (.xls) or Text (.txt) |
|
Example A803/ILT1700/Detector System Configurations by Application:
GERMICIDAL, HIGH INTENSITY, LOW PRESSURE MERCURY DETECTOR Specifically designed to measure high intensity low-pressure mercury lamps at 254 nm. System Configuration: A803, ILT1700, Detector: SED008/W, Range: 2 e-7 to 1 E+2 W/cm²
GERMICIDAL, LOW-PROFILE, HIGH-INTENSITY GERMICIDAL DETECTOR Filtered for measurement of 254 nm for narrow and broad spectrum lamps, in a low profile housing for versatility. System Configuration: A803, ILT1700, Detector: XRD140T254, Range: 7e-9 to 6e+0 W/cm²
PHOTODEGRADATION / PHOTOSTABILITY MAPPING UVA AND VISIBLE (WHITE) LIGHT TEST SYSTEM Measurement of UVA in W/m² and VIS in Lux for long term testing and datalogging during degradation studies. System Configuration: A803, ILT1700, Detectors: SED033/UVA/TD, 5e-6 to 5e+3 W/m² and SED033/Y/TD, 5.98e-2 to 2.09e+5 lux.
SOLAR MEASUREMENT, UVA and UVB For solar measurements of UVA and UVB from the sun or solar simulators. These sensors have been selected for the improved out-of-band rejection. Additional sensors for Broadband, Visible and IR regions are also available. System Configuration: **A803, ILT1700, Detectors: SED005/UVA/TD, 5e-9 to 4e+0 W/cm² andSED240/UVB/SPS300/T/W, 4e-8 to 2e-2 W/cm²
PHOTOTHERAPY, UVA and UVB For rapid testing of UV irradiance levels during patient exposure. System Configuration: **A803, ILT1700, Detectors: SED033/UVA/W, 3e-10 to 2e-1 W/cm² and SED240/UVB-1/W, 2e-7 to 6e-2 W/cm²
PLANT GROWTH / PHOTOBIOLOGY For measurement of Photosynthetically Active Radiation, Blue light, Red light and Far red light levels to test the affects on plant growth. System Configuration: A803, ILT1700, Detectors: SED033/PAR/W, 5e-10 to 5e-1 uE/cm²/sec; SED033/TBLU/W, 3e-10 to 2e-1 W/cm²; SED033/TRED/W, 2e-10 to 1e-1 W/cm²; SED033/TFRD/W, 1.47e-10 to 1.47e-1 |
|